Zpracování obrazů ultrazvukového zobrazovacího systému GE VingMed System FiVe / Image processing of data from the ultrasound system GE VingMed System FivePokorný, Pavel January 2013 (has links)
The master thesis is focused on the principle of ultrasonographs and their modes, and ways of denoising data obtained by ultrasonographs. This project will also concentrate on removing the speckle noise, because the speckle noise has the largest share of the noise in the measured data in ultrasonographs. The other main objective was to describe advanced methods of filtration, especially the modifications of the median filter, such as the modified hybrid median filter and modification of the anisotropic diffusion, namely generalizing Perona-Malik anisotropic diffusion using restricted dissimilarity functions. These two filters were tested on images with artificial noise and on data captured using the ultrasound system GE VingMed System FiVe. Also, the program for filtering and display filtered ultrasound data captured by GE VingMed System FiVe is described.
|