The master thesis deals with comparison of atomic fluorescence spektrometry and high resolution continuum source atomic absorption spektrometry for three methods generation of volatile compounds Antimony. In both methods of atomic antimony detection, it compares the most common chemical generation of volatile compounds (hydrides) with two alternative methods - electrochemical and UV - photochemical. The values of performance parameters for the determination of Sb(III) and Sb(V) were determined for all the above combinations. In the case of chemical generation, a surprisingly almost four times higher limit of detection of Sb(III) was found in connection with AFS detection than AAS detection. The final part was devoted to UV - photochemical vapor generation, with AAS detection for Sb(III) reaching limit of detection 4,96 ppb, for Sb(V) 8,63 ppb. Although UV - photochemical generation of volatile antimony compounds did not reach such performance parameters as chemical or electrochemical generation, it was observed that the sensitivity of antimony determination increased greatly when introducing oxygen into the apparatus. The interference study also found a significant positive effect of Fe(II) on the generation efficiency, and this modification partially persisted without further introduction of these...
Identifer | oai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:435877 |
Date | January 2020 |
Creators | Adámková, Dominika |
Contributors | Červený, Václav, Hraníček, Jakub |
Source Sets | Czech ETDs |
Language | Czech |
Detected Language | English |
Type | info:eu-repo/semantics/masterThesis |
Rights | info:eu-repo/semantics/restrictedAccess |
Page generated in 0.0018 seconds