Return to search

Within-day and between-day reliability of body composition by air-displacement plethysmography in the bod pod

The purpose of this study was to determine the reliability of within-day and between-day measurements of body volume (BV), thoracic gas volume (Vtg), and body fat percentages (%fat) taken in the BOD POD. Fifty subjects (25 men, 25 women) of varying ages (20-81 yrs.), races, and body composition (2.2-51.1% fat) were tested under ideal conditions set forth by the manufacturer at the same time of day (± 3 hours) on three separate days within two weeks. On each day, testing was repeated in succession until three trials were obtained which met the manufacturer's criteria for acceptable BV and Vtg. Repeated measures ANOVA revealed no significant mean differences between the three within-day trials or for the first trial between each of the three days with the exception of between-day BV where day 1 was found to be 260 ml greater than days 2 and 3. Coefficient of variation (SD/mean* 100) and intra-class correlations were determined for each variable both within the three trials of each day and between the first trials of each of the days. For BV, the CV for the 150 cases of within-day measurements was.1 ± .01% while the between-day CV was .4 +.3%. For both within and between day BV, r = .99. The CV and correlations for Vtg was 2.8 ± 2.4% and .99 and 4.1 ± 2.6% and .98 for within-day and between-day measures, respectively. The mean absolute difference between measured Vtg and the predicted value (Vpred) from age, height, and weight was 0.471 ± 0.392 L which would result in body fat estimates that differ by 1.3 ± 0.3%. Estimates of body fat from Vpred were within +2% fat of those derived from measured Vtg values in 80% of the trials. Body fat percentage produced a within-day CV of 2.8 + 4.6% with an r = .99 and a between-day CV of 3.8 ± 5.9% with an r =.99. However, with the exclusion of 3 subjects with body fat <7%, the within-day and between-day CV decreased to 2.0 ± 1.8% and 2.7 ± 2.1%, respectively. The difference between the first two trials of each day was < 2% fat in 93% of the cases, and all 150 cases resulted in fat within + 2% in the three trials. In conclusion, these results support that BOD POD measurements of BV, Vtg, and % fat are reliable both within and between days for a wide range of adults. Also, to obtain the most reliable measurement of % fat in the BP, measuring thoracic gas volume and averaging the values obtained from two trials which agree within + 2 % fat is recommended. / School of Physical Education

Identiferoai:union.ndltd.org:BSU/oai:cardinalscholar.bsu.edu:handle/186640
Date January 1999
CreatorsSparks, Jada L.
ContributorsKaminsky, Leonard A., 1955-
Source SetsBall State University
Detected LanguageEnglish
Format62 leaves : ill. ; 28 cm.
SourceVirtual Press

Page generated in 0.0022 seconds