Tissue equivalent proportional counters (TEPCs) are commonly used for radiation monitoring in areas where a mixture of neutron and photon radiations may be present, such as those commonly encountered in nuclear power plants. In such radiation fields, the dose rate of each component can vary drastically from extremely low to very high. Among these possible combinations of radiation fields with very different dose rates, a mixed field of an intense photon and a weak neutron dose component is the more commonly encountered.
This study describes the measurement of lineal energy spectra carried out with a 5.1 cm (2 inch) diameter spherical TEPC simulating a 2 μm diameter tissue site in low energy (33 – 330 keV neutrons) mixed photon-neutron fields with varying dose rates generated by the McMaster University 1.25 MV double stage Tandetron accelerator. The Tandetron accelerator facility was employed to produce neutrons using thick 7Li targets via the 7Li(p, n)7Be reaction. A continuous spectrum of neutrons is generated at any selected proton beam energy which is very narrow at beam energies very close to the threshold of the reaction 1.88 MeV and becomes wider as the proton beam energy moves further away from the threshold energy of the reaction. Dose rates which resulted in dead times as high as 75% for the data acquisition system were employed to study the effect of dose rate on the measured quality factors, microdosimetric averages (y ̅_f and y ̅_D)absorbed dose and dose equivalent. The dose rate at a given beam energy was varied by changing the accelerator beam current. A variety of mixed neutron gamma fields was generated using neutron beams with mean energies extending approximately from 33 keV to 330 keV with the 7Li target using proton beam energies ranging from 1.89 to 2.5 MeV. In direct beams, 478 keV photons which are produced in the 7Li target via inelastic scattering interaction 7Li(p, p'γ)7Li dominate the low LET component of the mixed field of radiation. When a 2 cm thick polyethylene moderator was inserted between the neutron producing target and the counter, the low LET component of the mixed radiation field also contained 2.20 MeV gamma rays originating from 1H(n, γ)2H capture interactions in the moderator.
We have observed that high dose rates due to both photons and neutrons in a mixed field of radiation result in pile up of pulses and distort the lineal energy spectrum measured under these conditions. The pile up effect and hence the distortion in the lineal energy spectrum becomes prominent with dose rates which result in dead times larger than 25% for the high LET radiation component. In intense neutron fields, which may amount to 75% dead time, a 50% or even larger increase in values for the measured microsdosimetric averages and the neutron quality factor was observed. This study demonstrates that moderate dose rates which do not result in dead times of more than 20-25% due to either of the component radiations or due to both components of mixed field radiation generate results which are acceptable for operational health physics mixed neutron-gamma radiation monitoring using tissue equivalent proportional counters. / UOIT
Identifer | oai:union.ndltd.org:LACETR/oai:collectionscanada.gc.ca:OOSHDU.10155/90 |
Date | 01 April 2010 |
Creators | Qashua, Nael |
Contributors | Waker, Anthony |
Source Sets | Library and Archives Canada ETDs Repository / Centre d'archives des thèses électroniques de Bibliothèque et Archives Canada |
Language | English |
Detected Language | English |
Type | Thesis |
Page generated in 0.0021 seconds