Return to search

A method of verification of the total treatment time for the APBI (Accelerated Partial Breast Irradiation) devices: CONTURA Multilumen Balloon and SAVI Applicator

A simple method to verify the total treatment time generated by the treatment planning system (TPS) when the CONTURA MLB or the SAVI applicator are used for APBI treatments has been developed. The method compares the time generated by the TPS to a predicted time, calculated by inserting parameters obtained from the TPS in equations generated in this Thesis. The equations were generated by linearly fitting data from clinical cases that had been treated using the Contura MLB or the SAVI applicator at the Lynn Cancer Institute of the Boca Raton Regional Hospital. The parameters used were the PTV coverage, Air Kerma Strength, Balloon Volume (Contura data fit) and Evaluation PTV (SAVI data fit). As an outcome of this research, it is recommended that the plan should be reevaluated when the percent difference between the generated and the predicted times exceeds 5% for the Contura MLB, or 10% for the SAVI. / by Andreas Kyriacou. / Thesis (M.S.)--Florida Atlantic University, 2011. / Includes bibliography. / Electronic reproduction. Boca Raton, Fla., 2011. Mode of access: World Wide Web.

Identiferoai:union.ndltd.org:fau.edu/oai:fau.digital.flvc.org:fau_3777
ContributorsKyriacou, Andreas, Charles E. Schmidt College of Science, Department of Physics
PublisherBoca Raton, Fla.
Source SetsFlorida Atlantic University
LanguageEnglish
Detected LanguageEnglish
TypeText, Electronic Thesis or Dissertation
Formatx, 48 p. : ill. (some col.), electronic
Rightshttp://rightsstatements.org/vocab/InC/1.0/

Page generated in 0.002 seconds