A multi-detector spectrometer was constructed to measure garmia-gamma angular correlations in 154Gd from the decay of 154Eu. Gamma-ray multipole mixing ratios were extracted from the data. A modified fast-slow coincidence electronics arrangement with selective routing of data was used. One Ge(Li) detector and three NaI detectors were used in the system for rapid data acquisition. "Real-plus-random" and "random" coincidences were accumulated simultaneously with detectors set at selected angles. Mixing ratios for the first transition in the cascades of 1274-123--keV (2--2+-0+), 1005-123-keV (3+-2+-0+), and 723-996-keV (2--2+-O+) were measured as '5(1274) = 0.02 0.02, '5(1005) = -28+10' '5(723) = 0.02 0.02.'The measured correlation for the 592-1005 keV cascade (2--3+-2+) was analyzed by searching a 157 by 157 array of values for '5(592) and 8(1005); agreement was tested using a Q2 criterion. The best agreement for the mixing ratio of the 592-keV transition was 8(592) = 0.08 0:22; for the 1005-keV transition 8(1005) = -9.5+7. The value 8(1005) _ -9.5+7 is taken to be the better value as a result of experimental considerations. The value for the 592-keV transition had not been previously measured.Ball State UniversityMuncie, IN 47306
Identifer | oai:union.ndltd.org:BSU/oai:cardinalscholar.bsu.edu:handle/180524 |
Date | 03 June 2011 |
Creators | Weeber, Winton L. |
Contributors | Ober, David R. |
Source Sets | Ball State University |
Detected Language | English |
Format | iv, 48 leaves : ill. ; 28 cm. |
Source | Virtual Press |
Page generated in 0.0018 seconds