This master's thesis is concerned with the basis of ultra-acoustics and it explains the application of Doppler effect in medicine as a diagnostic method used for measuring blood flow velocity in blood vessels. The thesis also presents Doppler systems used for diagnostic purposes. It describes the principles of these systems, shows accompanying flowcharts and explains the function of these systems. There is also included the system design of the whole ultrasonic blood flow velocity meter. The system has been designed for the specified parameters with continuous 4 MHz operating frequency, generated intensity of ultrasound 100 mW/cm^2 and diameter D-shaped transmitting transducer 8 mm. In the design of personal solution there are described all the function blocks of the ultrasonic meter. There is also described the choice of appropriate components and designed all the required supply voltages. All the function blocks of the system are experimentally verified. The system has been designed with acoustic intercept and the resultant velocity curves are displayed on software oscilloscope on PC.