MapCheck measurements for 50 retrospective patient’s treatment plans suggested that MapCheck could be effectively employed in routine patient specific quality assurance in M6 Cyberknife with beams delivered at different treatment angles. However, these measurements also suggested that for highly intensity modulated MLC plans, field segments of width < 8 mm should further be analyzed with a modified (-4%) correction factor. Results of MC simulations of the M6 Cyberknife using the EGSnrc program for 2-5 millions of incident particles in BEAMnrc and 10-20 millions in DOSXYZnrc have shown dose uncertainties within 2% for open fields from 7.6 x 7.7 mm2 to 100 x 100 mm2. Energy and corresponding FWHM were optimized by comparing with water phantom measurements at 800 mm SAD resulting to E = 7 MeV and FWHM = 2.2 mm. Good agreement of dose profiles (within 2%) and outputs (within 3%) were found between the MC simulations and water phantom measurements for the open fields. / Includes bibliography. / Thesis (M.S.)--Florida Atlantic University, 2018. / FAU Electronic Theses and Dissertations Collection
Identifer | oai:union.ndltd.org:fau.edu/oai:fau.digital.flvc.org:fau_40944 |
Contributors | Neupane, Taindra (author), Shang, Charles (Thesis advisor), Leventouri, Theodora (Thesis advisor), Florida Atlantic University (Degree grantor), Charles E. Schmidt College of Science, Department of Physics |
Publisher | Florida Atlantic University |
Source Sets | Florida Atlantic University |
Language | English |
Detected Language | English |
Type | Electronic Thesis or Dissertation, Text |
Format | 93 p., application/pdf |
Rights | Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder., http://rightsstatements.org/vocab/InC/1.0/ |
Page generated in 0.0048 seconds