Return to search

Comparison of intraoral and extraoral scanners on the accuracy of digital model articulation

Introduction: Orthodontists increasingly rely on digital models in clinical practice. The ability of modern scanners to articulate digital models must be scientifically evaluated. Methods:Twenty five digital articulated models were produced from four digital scanners in five experimental groups. The resulting inter-arch measurements were compared to the gold standard. An acceptable range of 0.5mm more or less than the gold standard was used for evaluation. Results: iTero® and iTero® Element yielded all acceptable inter-arch measurements. The 3M™ True Definition and Ortho Insight 3D® with Regisil® bite registration produced four of six acceptable inter-arch measurements. The Ortho Insight 3D® with Coprwax ™ bite registration yielded three of six acceptable inter-ach measurements. Conclusions: The iTero® and iTero® Element produced the most accurately articulated models. The 3M™ True Definition and Ortho Insight 3D® with Regisil® were the next most accurate. The Ortho Insight 3D® scanner with Coprwax ™ was the least accurate method tested.

Identiferoai:union.ndltd.org:vcu.edu/oai:scholarscompass.vcu.edu:etd-5985
Date01 January 2017
CreatorsPorter, Jason L
PublisherVCU Scholars Compass
Source SetsVirginia Commonwealth University
Detected LanguageEnglish
Typetext
Formatapplication/pdf
SourceTheses and Dissertations
Rights© Jason Porter

Page generated in 0.0023 seconds