Return to search

The preparation and operation of lithium drift germanium detectors

Lithium drifted germanium detectors have been prepared for use as high resolution gamma ray spectrometers. The fabrication procedure and the problems which can arise during preparation are discussed in detail. Using the techniques described, germanium detectors having the following characteristics were prepared.
Active Volume 1.0 cm³ 0.5 cm³ 2.0 cm³ 1.7 cm³
Total Resolution at 661 keV 5.0 KeV 4.0 KeV 4.0 KeV 2.9 KeV / Science, Faculty of / Physics and Astronomy, Department of / Graduate

Identiferoai:union.ndltd.org:UBC/oai:circle.library.ubc.ca:2429/36563
Date January 1966
CreatorsThompson, Albert Charles
PublisherUniversity of British Columbia
Source SetsUniversity of British Columbia
LanguageEnglish
Detected LanguageEnglish
TypeText, Thesis/Dissertation
RightsFor non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.