In the period immediately following the implantation of a cochlear implant electrode array within the cochlear environment, the power required to stimulate the auditory nerve at preset current levels increases. This rise is due to increases in electrode impedance which in turn is suggested to be a result of tissue growth around the electrode array. The foreign body response initiated by the immune system encapsulates the array in a matrix of fibrous tissue, separating the electrode array from the rest of the body. A second change in electrode impedance occurs with the onset of electrical stimulation. A transitory reduction in impedance has been recorded in animals and humans after stimulation of electrodes. Impedance returns to pre-stimulation levels following the cessation of stimulation. It was suggested that these changes in impedance with stimulation were also related to the tissue growth around the electrode array. A more thorough understanding of the interface was required to ascertain these concepts.
Identifer | oai:union.ndltd.org:ADTP/270044 |
Date | January 2006 |
Creators | Newbold, Carrie |
Source Sets | Australiasian Digital Theses Program |
Language | English |
Detected Language | English |
Rights | Restricted Access: Abstract and Citation Only |
Page generated in 0.0017 seconds