Previous issue date: 2018-08-16 / Este trabalho consiste no desenvolvimento de um Equipamento Médico Assistencial
(EMA) que realiza ablação hepática, por meio de radiofrequência, para tratamento de
pacientes com câncer no fígado. No processo de desenvolvimento, vários aspectos foram
considerados, tais como: compreensão do comportamento do sinal de saída do EMA por meio
de modelagens fenomenológicas (MF) e Identificação de Sistemas (IS), além do
comportamento físico do fígado frente a radiofrequência.
Para verificar o desempenho do equipamento desenvolvido, os testes de exatidão,
monotonicidade e avaliação da qualidade do sinal de saída por meio do Fator de Crista (FC)
foram realizados de acordo com a Norma da ABNT - NBR-IEC 60601-2-2 (2013).
Os resultados obtidos nos ensaios em laboratório mostraram que os parâmetros
avaliados no EMA desenvolvido apresentaram valores melhores que os recomendados pela
Norma da ABNT - NBR-IEC 60601-2-2 (2013). Adicionalmente, as funções de transferências
obtidas, representaram com alto grau de correlação o comportamento do equipamento. / This work consists of the development of a Medical Assistance Equipment (MAE) that
performs hepatic ablation, by means of radiofrequency, for the treatment of patients with liver
cancer. In the development process, several aspects were considered, such as: understanding
the behavior of the EMA output signal through phenomenological modeling (MF) and System
Identification (IS), in addition to the physical behavior of the liver versus radiofrequency.
In order to verify the performance of the developed equipment, the tests of accuracy,
monotonicity and evaluation of the quality of the output signal through the Crest Factor (FC)
were performed according to the ABNT - NBR - IEC 60601-2-2 (2013).
The results obtained in the laboratory tests showed that the parameters evaluated in the
developed EMA presented better values than those recommended by the Standard of ABNT -
NBR-IEC 60601-2-2 (2013). In addition, the transfer functions obtained represented a high
degree of correlation with the behavior of the equipment.
Identifer | oai:union.ndltd.org:IBICT/oai:repositorio.unb.br:10482/32495 |
Date | 26 February 2018 |
Creators | Cavalcante, Gilvandson Costa |
Contributors | Jacobi, Ricardo Pezzuol, Rosa, Suélia de Siqueira Rodrigues Fleury |
Source Sets | IBICT Brazilian ETDs |
Language | Portuguese |
Detected Language | Portuguese |
Type | info:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/masterThesis |
Source | reponame:Repositório Institucional da UnB, instname:Universidade de Brasília, instacron:UNB |
Rights | A concessão da licença deste item refere-se ao termo de autorização impresso assinado pelo autor com as seguintes condições: Na qualidade de titular dos direitos de autor da publicação, autorizo a Universidade de Brasília e o IBICT a disponibilizar por meio dos sites www.bce.unb.br, www.ibict.br, http://hercules.vtls.com/cgi-bin/ndltd/chameleon?lng=pt&skin=ndltd sem ressarcimento dos direitos autorais, de acordo com a Lei nº 9610/98, o texto integral da obra disponibilizada, conforme permissões assinaladas, para fins de leitura, impressão e/ou download, a título de divulgação da produção científica brasileira, a partir desta data., info:eu-repo/semantics/openAccess |
Page generated in 0.0015 seconds