Frequency domain functional near infrared spectroscopy (FD-fNIRS) is a tissue optical measurement technique used to measure absolute haemoglobin concentrations in brain tissue. This work is intended to be the first step in the development of a wearable, low-cost FD-fNIRS device for neurofeedback applications. The system requirements were generated from a review of relevant literature. A simplified system architecturewas developed based on the various instrumentation methodologies proposed by various authors. The functional blocks of this system were prototyped and their performance was evaluated. The developed vertical-cavity surface-emitting laser (VCSEL) current source was found to have a span of 10uA which meets the design specifications. Challenges exist in optimally biasing silicon photomultiplier (SiPM), which is susceptible to optical and electronic noise sources.
Identifer | oai:union.ndltd.org:UPSALLA1/oai:DiVA.org:kth-319211 |
Date | January 2022 |
Creators | Nareshkumar, Rohit Rathnam |
Publisher | KTH, Skolan för kemi, bioteknologi och hälsa (CBH) |
Source Sets | DiVA Archive at Upsalla University |
Language | English |
Detected Language | English |
Type | Student thesis, info:eu-repo/semantics/bachelorThesis, text |
Format | application/pdf |
Rights | info:eu-repo/semantics/openAccess |
Relation | TRITA-CBH-GRU ; 2022:278 |