Return to search

Feature Based Registration of Ultrasound and CT Data of a Scaphoid

Computer assisted surgery uses a collection of different techniques including but not limited to: CT-guided, fluoroscopy-guided, and ultrasound-guided imaging which allows medical staff to view bony anatomy of a patient in relation to surgical tools on a computer screen. By providing this visual data to surgeons less invasive surgeries can be performed on a patient's fractured scaphoid. The data required for a surgeon to perform a minimally invasive surgery while looking only at a computer screen, and not directly at a patient's anatomy, will be provided by CT and ultrasound data. We will discuss how ultrasound and CT data can be used together to allow a minimally invasive surgery of the scaphoid to be performed.

In this thesis we will explore two techniques of registering segmented ultrasound images to CT data; an Iterative Closest Point (ICP) approach, and an Unscented Kalman Filter-based Registration (UKF). We use two different ultrasound segmentation methods; a semi-automatic segmentation, and a Bayesian segmentation technique. The segmented ultrasound data is then registered to a CT volume. The success or failure of the
registrations is measured by the error calculated in mapping the corresponding land-
marks to one another and calculating the target registration error. The results show that the Unscented Kalman Filter-based registration using the Bayesian segmentation of ultrasound images has the least registration error, and has the most robustness to error in initial alignment of the two data sets. / Thesis (Master, Computing) -- Queen's University, 2010-05-28 11:17:31.934

Identiferoai:union.ndltd.org:LACETR/oai:collectionscanada.gc.ca:OKQ.1974/5689
Date28 May 2010
CreatorsKoslowski, Brian
ContributorsQueen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))
Source SetsLibrary and Archives Canada ETDs Repository / Centre d'archives des thèses électroniques de Bibliothèque et Archives Canada