Antimony film electrode was studied for the use in a voltammetric analysis of organic compounds. The substance chosen as an analyte was trifluralin, which is used as a pesticide. The comparison of different substrate electrodes was carried out between five electrodes, which were gold, silver, copper, polished amalgam and glassy carbon electrode (GCE). Best performance was observed on antimony film glassy carbon electrode (SbFGCE). It provided higher sensitivity and lower limit of quantification in comparison with bare GCE. The antimony film was stable and it provided good reproducibility (RSD = 5.2 %). Parameters of an electrochemical preparation of SbFGCE were optimized. Conditions for determination of concentration of trifluralin were optimized on newly prepared SbFGCE. The best conditions were in a solution of methanol and 0.1 M hydrochloric acid in 1:1 ratio measured by differential pulse voltammetry. The limit of quantification was determined as 1.2·10-6 mol·l-1 . A direct voltammetric measurement on SbFGCE was carried out in a model river sample. Lower limits of quantification were achieved with solid phase extraction (SPE). Recovery values were 86 ± 8 % in deionized water with a preconcentration factor of 125. The limit of quantification was lowered to value 1.1·10-8 mol·l-1 . The extraction...