This master thesis evaluated the performance of a micro-CT system consisting of Hamamatsu microfocus X-ray source L10951-04 and CMOS flat panel C7942CA-22. The X-ray source and flat panel have been characterised in terms of dark current, image noise and beam profile. Additionally, the micro-CT system’s spatial resolution, detector lag and detector X-ray response have been measured. Guidance for full image correction and methods for characterisation and performance test of the X-ray source and detector is presented. A spatial resolution of 7 lp/mm at 10 % MTF was measured. A detector lag of 0.3 % was observed after ten minutes of radiation exposure. The performance of the micro-CT system was found to be sufficient for high resolution X-ray imaging. However, the detector lag effect is strong enough to reduce image quality during subsequent image acquisition and must either be avoided or corrected for.
Identifer | oai:union.ndltd.org:UPSALLA1/oai:DiVA.org:kth-155454 |
Date | January 2014 |
Creators | Baumann, Michael |
Publisher | KTH, Skolan för teknik och hälsa (STH) |
Source Sets | DiVA Archive at Upsalla University |
Language | English |
Detected Language | English |
Type | Student thesis, info:eu-repo/semantics/bachelorThesis, text |
Format | application/pdf |
Rights | info:eu-repo/semantics/openAccess |
Relation | TRITA-STH ; 2014:87 |
Page generated in 0.0019 seconds