General testing method for refractive surfaces based on reverse Hartmann test

The testing technique with high dynamic range is required to meet the measurement of refractive wavefront with large distortion from test refractive surface. A general deflectometric method based on reverse Hartmann test is proposed to test refractive surfaces. Ray tracing of the modeled testing system is performed to reconstruct the refractive wavefront from test surface, in which computer-aided optimization of system geometry is performed to calibrate the geometrical error. For the refractive wavefront error with RMS 255 mu m, the testing precision better than 0.5 mu m is achieved.

Identiferoai:union.ndltd.org:arizona.edu/oai:arizona.openrepository.com:10150/627174
Date23 August 2017
CreatorsWang, Daodang, Xu, Ping, Liang, Rongguang, Ming, Kong, Zhao, Jun, Gong, Zhidong, Mo, Linhai, Mo, Shuhui, Xie, Zhongmin
ContributorsUniv Arizona, Coll Opt Sci
PublisherSPIE-INT SOC OPTICAL ENGINEERING
Source SetsUniversity of Arizona
LanguageEnglish
Detected LanguageEnglish
TypeArticle
Rights© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE).
Relationhttps://www.spiedigitallibrary.org/conference-proceedings-of-spie/10373/2273189/General-testing-method-for-refractive-surfaces-based-on-reverse-Hartmann/10.1117/12.2273189.full