An automated probe for thermal conductivity measurements

A transient technique was validated for making thermal conductivity measurements. The technique incorporated a small, effectively spherical, heat source and temperature sensing probe. The actual thermal conductivity measurements lasted 30 seconds. After approximately 15 minutes of data reduction, a value for thermal conductivity was obtained. The probe yielded local thermal conductivity measurements. Spherical sample volumes less than 8 cm² were required for the materials tested. Thermal conductivity (and moisture) distributions can be measured for relatively dry or wetted samples.

The technique employs an encapsulated bead thermistor. A thermistor, more commonly used as a temperature transducer, has the inherent feature of being readily self-heated. A computer-based data acquisition and control system regulates the power supplied to the thermistor such that its self-heated temperature response approximates a step change. Thermal conductivity is deduced from the transient measurement of the power dissipated by the probe as a function of time.

The technique was used to measure the thermal conductivity of fifteen liquids and five insulation materials. Two different thermistor types, glass-encapsulated and Teflon-encapsulated, were evaluated. Capabilities and limitations of each probe type and the measurement technique, in general, were observed. / M.S.

Identiferoai:union.ndltd.org:VTETD/oai:vtechworks.lib.vt.edu:10919/101183
Date January 1987
CreatorsDougherty, Brian P.
ContributorsMechanical Engineering
PublisherVirginia Polytechnic Institute and State University
Source SetsVirginia Tech Theses and Dissertation
LanguageEnglish
Detected LanguageEnglish
TypeThesis, Text
Formatxiii, 118 leaves, application/pdf, application/pdf
RightsIn Copyright, http://rightsstatements.org/vocab/InC/1.0/
RelationOCLC# 17605183