Point-of-care creatinine testing for kidney function measurement prior to contrast-enhanced diagnostic imaging: evaluation of the performance of three systems for clinical utility

Yes / Acute kidney injury (AKI) can occur rarely in patients exposed to iodinated contrast and result in contrast-induced AKI (CI-AKI). A key risk factor is the presence of pre-existing chronic kidney disease (CKD), therefore it is important to assess patient risk and obtain kidney function measurement prior to administration. Point of care (PoC) testing provides an alternative strategy but there remains uncertainty, with respect to diagnostic accuracy and clinical utility.
A device study compared three PoC analysers (Nova StatSensor, Abbott i-STAT, Radiometer ABL800 FLEX) with a reference laboratory standard (Roche Cobas 8000 series, enzymatic creatinine). Three hundred adult patients attending a UK hospital phlebotomy department were recruited to have additional blood samples for analysis on the PoC devices.
The ABL800 FLEX had the strongest concordance with laboratory measured serum creatinine (mean bias=-0.86, 95% limits of agreement = -9.6 to 7.9) followed by the i-STAT (average bias=3.88, 95% limits of agreement = -8.8 to 16.6) and StatSensor (average bias=3.56, 95% limits of agreement = -27.7 to 34.8). In risk classification, the ABL800 FLEX and i-STAT identified all patients with an eGFR≤30, whereas the StatSensor resulted in a small number of missed high-risk cases (n=4/13) and also operated outside of the established performance goals.
The screening of patients at risk of CI-AKI may be feasible with PoC technology. However in this study it was identified that the analyser concordance with the laboratory reference varies. It is proposed that further research exploring PoC implementation in imaging department pathways is needed. / Yorkshire and Humber Academic Health Science Network (Grant Number: YHP0318)

Identiferoai:union.ndltd.org:BRADFORD/oai:bradscholars.brad.ac.uk:10454/15685
Date19 April 2018
CreatorsSnaith, Beverly, Harris, Martine A., Shinkins, B., Jordaan, M., Messenger, M., Lewington, A.
Source SetsBradford Scholars
LanguageEnglish
Detected LanguageEnglish
TypeArticle, Accepted manuscript
Rights© 2018 Walter de Gruyter GmbH. Full-text reproduced in accordance with the publisher's self-archiving policy., Unspecified

Page generated in 0.0179 seconds