Return to search

Improvements to the kidney dialysis machine

Five changes were suggested to improve the kidney dialysis machine, including a conductivity control feedback system, a positive pressure deaeration system, a balancing system for dialysate flow control, a diaphragm-type blood pump, and an ultrafiltration control system using blood pressure measurement. A kidney dialysis machine was constructed to test the changes for possible use in future dialysis machine designs. Tests were made on each of the five system improvements. All five ideas had some merit for a commercial dialysis machine. A description of the investigation and the various factors used in the development and testing of the five systems is included. / Master of Science

Identiferoai:union.ndltd.org:VTETD/oai:vtechworks.lib.vt.edu:10919/41807
Date30 March 2010
CreatorsHundley, Robert Wynne
ContributorsMechanical Engineering, Hurst, Charles J., Reinholtz, Charles F., Moses, Hal L.
PublisherVirginia Tech
Source SetsVirginia Tech Theses and Dissertation
LanguageEnglish
Detected LanguageEnglish
TypeThesis, Text
Formatvi, 79 leaves, BTD, application/pdf, application/pdf
RightsIn Copyright, http://rightsstatements.org/vocab/InC/1.0/
RelationOCLC# 26088312, LD5655.V855_1992.H862.pdf

Page generated in 0.0021 seconds