Diploma thesis presents the description of laser induced breakdown spectroscopy (LIBS) analytical method. The possibility to decrease LIBS detection limits using double pulse LIBS (DP-LIBS) is also detailed. Properties of laser induced plasma (LIP) are discussed. Thesis describes the optimisation process of DP-LIBS setup at laser spectroscopy laboratory of Faculty of Mechanical Engineering, Brno University of Technology. Optimisation was carried out for detection of chrome in steel with respect to detection limits and ablation crater size. Optimised DP-LIBS setup is appropriate for mapping chrome distribution in steel samples.
Identifer | oai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:229113 |
Date | January 2010 |
Creators | Roščák, Michal |
Contributors | Novotný, Karel, Kaiser, Jozef |
Publisher | Vysoké učení technické v Brně. Fakulta strojního inženýrství |
Source Sets | Czech ETDs |
Language | Slovak |
Detected Language | English |
Type | info:eu-repo/semantics/masterThesis |
Rights | info:eu-repo/semantics/restrictedAccess |