The preclinical imaging method micro-CT (microtomography) allows the visualization and quantification of the structure of samples at a resolution of micrometers. Its' importance is increasing globally. In addition to several advantages (non-destructive, the possibility of direct 3D analysis, time efficiency, etc.), micro-CT also has some significant limitations (problematic validation of results, image artifacts, significant influence of image modifications, etc.). This thesis focuses on the application of micro-CT in the field of research and development of metallic and non-metallic materials promoting bone healing with their possible clinical applications. The first part addresses the limitations of micro-CT through several studies. A comparison of pore sizes in biomaterials utilizing scanning electron microscopy (SEM) and micro-CT was performed, and the complications of pore size evaluation were presented. SEM image analysis leads to significantly higher values than micro-CT (approximately three times), which allows for comparison of the studies using only one of these methods. Validation of micro-CT 3D analysis results based on calibration phantoms with complex structure, to date, is not possible. We therefore developed software generating phantom datasets of 3D objects with well-defined...