Objectives: Resolution recovery algorithms (IR-RR) were recently proposed as tools to improve quality of SPECT images through better resolution. The aim of the presented study was to investigate the effect of IR-RR on myocardial perfusion SPECT studies. Methods: Phantom and clinical studies were performed using SPECT-CT Infinia/Hawkeye (GE Healthcare). NEMA triple line phantom was scanned according to NEMA procedures. Cold sphere and cardiac phantom were scanned under clinical conditions (900 -angled detectors, 60 views and circular orbit) and reconstructed via IR-RR (Evolution for Cardiac, GE, 12 iterations and 10 subsets), OSEM (ordered subset expectation maximization, 2 iterations and 10 subsets) and FBP (filtered back projection). IR-RR and OSEM reconstructions were used with/without attenuation and scatter correction (ACSC). The effect of post-reconstruction filtering was evaluated. In clinical studies two data-sets were used (half-time and full-time). Conventional scan using full-time (20 sec per stress projection and 25 sec per rest projection) was reconstructed via FBP and IR-RR; half-time scan (10 sec stress, 12 sec rest studies) was reconstructed via IR-RR. End-diastolic volume (EDV), end-systolic volume (ESV) and left ventricular ejection fraction (EF) were calculated using two software...
Identifer | oai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:322225 |
Date | January 2013 |
Creators | Skibová, Daniela |
Contributors | Šámal, Martin, Judas, Libor, Zimák, Jaroslav |
Source Sets | Czech ETDs |
Language | Czech |
Detected Language | English |
Type | info:eu-repo/semantics/doctoralThesis |
Rights | info:eu-repo/semantics/restrictedAccess |
Page generated in 0.0018 seconds