Return to search

High Frequency Pure Tone Audiometry and High Frequency Distortion Product Otoacoustic Emissions: A Correlational Analysis

Previous studies show that pure tone thresholds are strongly correlated with distortion product otoacoustic emission amplitudes when evaluating the frequency range from 1 to 8 kHz (Avan & Bonfils, 1993). Little is known about correlations between these two measures at higher frequencies from 9-16 kHz. This study compared pure tone thresholds and distortion product otoacoustic emissions (DPOAEs) in this high frequency range for 29 normal hearing subjects ages 18-30. Pure tone thresholds were obtained at 250-16 kHz and distortion product otoacoustic emissions (DPOAE) 2,211-17,675 were measured in the same ears. DPOAE amplitudes were measured using a constant F1/F2 ratio of 1.2, with F2 values ranging from 2,211-17,675 Hz. Data obtained from 50 ears showed a decline in DPOAE amplitude with increasing frequency of the F1 and F2 primary stimulus tones. Behavioral thresholds demonstrated an increase with increasing frequency of the pure tone stimulus. Pearson r-correlation analysis demonstrated a weak relationship between measures. Further investigation revealed that equipment variables prevented accurate readings.

Identiferoai:union.ndltd.org:pdx.edu/oai:pdxscholar.library.pdx.edu:open_access_etds-2687
Date01 January 2003
CreatorsLavoie, Kimberly J.
PublisherPDXScholar
Source SetsPortland State University
Detected LanguageEnglish
Typetext
Formatapplication/pdf
SourceDissertations and Theses

Page generated in 0.0194 seconds