Developing a method for insertion of soft neural probes into peripheral nerves

The main objective of this project was to develop a method for soft neural probe insertion into a nerve. These soft probes are made of a silicon elastomer that is a few orders of magnitude less stiff than the nerve, which makes the insertion process particularly challenging. To overcome this challenge a tungsten microwire was used as an insertion shuttle to help penetrate the nerve by increasing the overall stiffness of the probe. At a first stage, the insertion process was tested on a nerve phantom (e.g. agarose gel with PDMS membrane) to control the insertion parameters and validate the insertion platform. Once the envisioned insertion method was established, the probe was implanted in a real rodent nerve. / <p>Examensarbetet är utfört vid Institutionen för teknik och naturvetenskap (ITN) vid Tekniska fakulteten, Linköpings universitet</p>

Identiferoai:union.ndltd.org:UPSALLA1/oai:DiVA.org:liu-186667
Date January 2022
CreatorsMelander, Klara
PublisherLinköpings universitet, Laboratoriet för organisk elektronik, Linköpings universitet, Tekniska fakulteten
Source SetsDiVA Archive at Upsalla University
LanguageEnglish
Detected LanguageEnglish
TypeStudent thesis, info:eu-repo/semantics/bachelorThesis, text
Formatapplication/pdf
Rightsinfo:eu-repo/semantics/openAccess