Master of Science / Department of Biological & Agricultural Engineering / Mei He / Anemia affects about 25% of the world’s population and causes roughly 8% of all disability cases. The development of an affordable point-of-care (POC) device for detecting anemia could be a significant for individuals in underdeveloped countries trying to manage their anemia. The objective of this study was to design and fabricate a 3D printed, low cost microfluidic mixing chip that could be used for the diagnosis of anemia.
Microfluidic mixing chips use capillary flow to move fluids without the aid of external power. With new developments in 3D printing technology, microfluidic devices can be fabricated quickly and inexpensively. This study designed and demonstrated a passive microfluidic mixing chip that used capillary force to mix blood and a hemoglobin detecting assay.
A 3D computational fluid dynamic simulation model of the chip design showed 96% efficiency when mixing two fluids. The mixing chip was fabricated using a desktop 3D printer in one hour for less than $0.50. Blood samples used for the clinical validation were provided by The University of Kansas Medical Center Biospecimen Repository. During clinical validation, RGB (red, green, blue) values of the hemoglobin detection assay color change within the chip showed consistent and repeatable results, indicating the chip design works efficiently as a passive mixing device. The anemia detection assay tended to overestimate hemoglobin levels at lower values while underestimating them in higher values, showing the assay needs to go through more troubleshooting.
Identifer | oai:union.ndltd.org:KSU/oai:krex.k-state.edu:2097/32875 |
Date | January 1900 |
Creators | Plevniak, Kimberly |
Publisher | Kansas State University |
Source Sets | K-State Research Exchange |
Language | en_US |
Detected Language | English |
Type | Thesis |