The main objective of this project is to design, construct and validate a portable prototype of a device that is capable of performing a test to accurately measure Capillary Refilling Time (CRT), and to analyze the results with defined parameters; force, area, pressure (compression) and time. This prototype is dedicated to study and evaluate CRT readouts for different pressure values, collected from healthy subjects.The presented prototype of this study is capable of producing skin compressing and to measure the refilling time of capillaries following this compression. This prototype introduces accuracy, mechanical reproducibility and controlling options for the applied pressure and compression time. The presented prototype is non-invasive, portable and it can be used to conduct more CRT tests and other capillary refilling studies.CRT measurement is done by calculating time interval starting from the first point when the applied pressure is released; ending with the recording point at the time when the concentration of red blood cells has reached the level of its pre-occlusion values.Based on the calculated CRT values and the number of iterations of the test in CRT tables, one can observe that given the same applied pressure value, CRT values do not significantly vary when the test is repetitively conducted on the same subject and on the same site.
Identifer | oai:union.ndltd.org:UPSALLA1/oai:DiVA.org:liu-81496 |
Date | January 2012 |
Creators | Shamsudin, Nebil |
Publisher | Linköpings universitet, Biomedicinsk instrumentteknik, Linköpings universitet, Tekniska högskolan |
Source Sets | DiVA Archive at Upsalla University |
Language | English |
Detected Language | English |
Type | Student thesis, info:eu-repo/semantics/bachelorThesis, text |
Format | application/pdf |
Rights | info:eu-repo/semantics/openAccess |
Page generated in 0.0019 seconds