This thesis studies comparison and connection of two spectrometric methods – low energy ion scattering spektrometry (LEIS) and secondary ion mass spectrometry (SIMS). SIMS method, despite its many positive qualities, suffers of so called matrix effect, which makes quantifiaction of data very difficult. LEIS method on the other hand is immune to this effect and so it’s suitable completion of SIMS method. As a convenient sample have been chosen AlGaN samples with various concentration of gallium and aluminium. In the first part of thesis is introduced physical essence of SIMS and LEIS method, experimental details and studied samples. In second part of the thesis there’s a description of measurements and comparison of data gained by each method.
Identifer | oai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:402575 |
Date | January 2019 |
Creators | Staněk, Jan |
Contributors | Šik, Ondřej, Bábor, Petr |
Publisher | Vysoké učení technické v Brně. Fakulta strojního inženýrství |
Source Sets | Czech ETDs |
Language | Czech |
Detected Language | English |
Type | info:eu-repo/semantics/masterThesis |
Rights | info:eu-repo/semantics/restrictedAccess |
Page generated in 0.0023 seconds