Return to search

Axial alignment in a ring-collection beta-ray spectrometer

A thin-lens beta-ray spectrometer using ring-focus collection was modified. These modifications consisted of; 1) a centering mechanism enabling the source-detector axis to be aligned with the magnetic axis; 2) an extension of the vacuum chamber placing the detector further from the magnet coils. The latter considerably decreased the magnetic shielding
requirements for the detector. A misalignment of 0.25 mm. for parallel axes and of 0°09' for intersecting axes produced noticeably poorer performance. Using a gathering power of .70%, a resolving power of .94% was obtained for the 661.6 Kev. K-conversion peak of Cs ¹³⁷. / Science, Faculty of / Physics and Astronomy, Department of / Graduate

Identiferoai:union.ndltd.org:UBC/oai:circle.library.ubc.ca:2429/40123
Date January 1960
CreatorsEarle, Eric Davis
PublisherUniversity of British Columbia
Source SetsUniversity of British Columbia
LanguageEnglish
Detected LanguageEnglish
TypeText, Thesis/Dissertation
RightsFor non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.