A thin-lens beta-ray spectrometer using ring-focus collection was modified. These modifications consisted of; 1) a centering mechanism enabling the source-detector axis to be aligned with the magnetic axis; 2) an extension of the vacuum chamber placing the detector further from the magnet coils. The latter considerably decreased the magnetic shielding
requirements for the detector. A misalignment of 0.25 mm. for parallel axes and of 0°09' for intersecting axes produced noticeably poorer performance. Using a gathering power of .70%, a resolving power of .94% was obtained for the 661.6 Kev. K-conversion peak of Cs ¹³⁷. / Science, Faculty of / Physics and Astronomy, Department of / Graduate
Identifer | oai:union.ndltd.org:UBC/oai:circle.library.ubc.ca:2429/40123 |
Date | January 1960 |
Creators | Earle, Eric Davis |
Publisher | University of British Columbia |
Source Sets | University of British Columbia |
Language | English |
Detected Language | English |
Type | Text, Thesis/Dissertation |
Rights | For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use. |