A T wave peak and offset detector based on an unpublished lead transformation that can be briefly described as multilead linear regression was proposed and implemented afterwards. Potential of the transformation as a useful QRS detection tool was revealed later on. Proposed QRS detector was put to the test of CSE database. Results were compared with work of other authors. Results of T wave peak and offset detector were introduced in visual way.
Identifer | oai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:220848 |
Date | January 2014 |
Creators | Brandejs, Jakub |
Contributors | Halámek, Josef, Veselý, Petr |
Publisher | Vysoké učení technické v Brně. Fakulta elektrotechniky a komunikačních technologií |
Source Sets | Czech ETDs |
Language | Czech |
Detected Language | English |
Type | info:eu-repo/semantics/masterThesis |
Rights | info:eu-repo/semantics/restrictedAccess |
Page generated in 0.0018 seconds