Return to search

Stanovení standardní nejistoty při konfirmaci teplotní komory / Evaluation of standard uncertainty of Climatic Chamber

The thesis deals with the problematics of confirmation of the climatic chamber CTS 65/50, which is a component of CVVOZE testing laboratory. Theoretical part deals with the analysis and description of methods of confirmation of climatic chambers based primarily on technical standards. This part then contains a brief description of a theory of measurement uncertainty. Practical part deals with the design of measurement procedure, which was used for confirmation measurement of ten temperature setpoints. This section then also contains an analysis of the measured data with the evaluation of standard combined uncertainty of temperature measurement. The last part contains an evaluation of the confirmation results, which were the basis for the determination of whether the climatic chamber meets the requirements for the steady-state temperature stability inside the workspace resulting from the standards for the climatic testing

Identiferoai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:377087
Date January 2018
CreatorsDojava, Václav
ContributorsHavránek, Zdeněk, Šedivá, Soňa
PublisherVysoké učení technické v Brně. Fakulta elektrotechniky a komunikačních technologií
Source SetsCzech ETDs
LanguageCzech
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/masterThesis
Rightsinfo:eu-repo/semantics/restrictedAccess

Page generated in 0.002 seconds