Return to search

Spolehlivostní analýza keramické hlavice totální kyčelní endoprotézy / Reliability analysis of the hip joint endoprosthesis ceramic head

This diploma thesis deals with probability of fracture ceramic head of total hip endoprosthesis during tests by the standard ISO 7206-5. In the first part is made stress analysis with method of finite elements in program ANSYS. Six types of ceramic head with different geometry was analysed for ideal geometry of head and shaft, and also for shape imperfections from nominal conical shaft and head. The final values of principal stress from stress analysis were entry values for the second part – probability of fracture analysis. Weibull weakest-link theory was applied, namely 3-Parameter and 2-Parameter Weibull distribution.

Identiferoai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:228919
Date January 2010
CreatorsMálek, Michal
ContributorsJaníček, Přemysl, Fuis, Vladimír
PublisherVysoké učení technické v Brně. Fakulta strojního inženýrství
Source SetsCzech ETDs
LanguageCzech
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/masterThesis
Rightsinfo:eu-repo/semantics/restrictedAccess

Page generated in 0.0432 seconds