Return to search

Mechanická analýza vlivu výrobních odchylek na styk hlavice a jamky totální endoprotézy kyčelního kloubu / Mechanical Analysis of the Influence of Fabrication Tolerances on Contact Conditions between the Acetabular Cup and the Femoral Head in Total Hip Replacement

In endoprosthesis surgery there are typically a high percentage of implant defects, these can lead to failure of the whole prosthesis. One type of total hip replacement function loss is acetabular cup loosening from the pelvic bone. This disertation examines manufacture perturbations as one of the possible reasons for this kind of failure. Both dimension and geometry manufacturing perturbations of ceramic head and polyethylen cup were analyzed. We find that perturbations in the variables analysed here affect considered values of contact pressure and frictional moment. Furthermore, contact pressure and frictional moment are quantities affecting replacement success and durability. From obtained results it can be recommended to fit head and cup with a clearance of between 0 mm and 0.05 mm. It can not be recommend using interference type of fit because of strong deterioration of the contact conditions. Roundness perturbation of ceramic head should not exceed 0.025 mm.

Identiferoai:union.ndltd.org:nusl.cz/oai:invenio.nusl.cz:233902
Date January 2010
CreatorsKoukal, Milan
ContributorsFlorian, Zdeněk
PublisherVysoké učení technické v Brně. Fakulta strojního inženýrství
Source SetsCzech ETDs
LanguageCzech
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/doctoralThesis
Rightsinfo:eu-repo/semantics/restrictedAccess

Page generated in 0.002 seconds