Milk Protein Analysis by Automated Ultraviolet Spectroscopy

An automated continuous flow analysis for milk protein utilizing ultraviolet spectroscopy resulted in a correlation coefficient of 0.972 and a standard estimate of error of 0.082 percent protein when 30 samples were compared with the acid orange 12 dye binding method. Milk was solubilized in 95-105 volumes of acetic acid using the Technicon Auto Analyzer II and measured for absorbance at 274 nanometers. No deviation from the standard method due to mastitis and varying milk fat concentration was observed. A theoretical model indicated a possible standard estimate of error 0.059 percent protein due to inter- and intra-breed variation in milk protein fraction distribution. Potassium dichromate and other preservatives interfered with the results.

Identiferoai:union.ndltd.org:UTAHS/oai:digitalcommons.usu.edu:etd-6190
Date01 May 1975
CreatorsWilkinson, R. Ford
PublisherDigitalCommons@USU
Source SetsUtah State University
Detected LanguageEnglish
Typetext
Formatapplication/pdf
SourceAll Graduate Theses and Dissertations
RightsCopyright for this work is held by the author. Transmission or reproduction of materials protected by copyright beyond that allowed by fair use requires the written permission of the copyright owners. Works not in the public domain cannot be commercially exploited without permission of the copyright owner. Responsibility for any use rests exclusively with the user. For more information contact digitalcommons@usu.edu.