Return to search

A Technical and Clinical Assessment of Stereotactic Registration Techniques to Improve MRI Guided Needle Navigation in Prostate Cancer Targeting

Prostate cancer is prevalent among men and one of the few cancer sites where local therapies currently target the entire organ instead of tumour. MRI holds promise in accurately depicting regions of cancer burden within the prostate gland and guiding tumour-targeted diagnostics and therapeutics. The clinical performance of a novel stereotactic MRI-guided needle navigation system for prostate cancer targeting was evaluated. Mean absolute in-plane stereotactic needle-targeting error for 10 patients was 2.2 mm and mean absolute depth error was 6.5 mm, highlighting a need to improve technical accuracy of the system. Consequently, alternative stereotactic registration techniques were investigated. Metrics of performance were in-plane stereotactic needle-targeting error, depth error, and registration time. A Z-shaped fiducial motif using automated registration performed best in phantom experiments with an in-plane error of 2.0 mm and depth error of 1.0 mm. These results will guide further software and hardware development to improve clinical performance.

Identiferoai:union.ndltd.org:LACETR/oai:collectionscanada.gc.ca:OTU.1807/18886
Date15 February 2010
CreatorsSuljendic, Denis
ContributorsMenard, Cynthia
Source SetsLibrary and Archives Canada ETDs Repository / Centre d'archives des thèses électroniques de Bibliothèque et Archives Canada
Languageen_ca
Detected LanguageEnglish
TypeThesis

Page generated in 0.0019 seconds