The thesis deals with the method for the improvement of the measurement accuracy in the optical frequency spectra measurements based on tunable optical filters. The thesis starts with the selection of the proper optical filter that is suitable for accurate light spectrum measurement. The selected filter was then used for the design and realization of the measurement system for the inspection of the fibre Bragg gratings. The system incorporates a reference block for the compensation of environmental influences, an interferometric verification subsystem and a PC - based control software implemented in LabView. The preliminary experimental verification of the measurement principle and the measurement system functionality were carried out on a testing rig with a specially prepared concrete console in the UJV Řež. Advanced tests were conducted on nuclear power plant’s containment shape deformation measurement in Temelin.