The work describes the design and manufacturing of a test device for post-mortem measurements of electrodes of electrochemical cells using X-ray diffraction spectroscopy. The theoretical part describes the diffractometer Rigaku Miniflex 600, for which the product is intended. At the same time, an overview of solutions in various applications is processed here in the form of a recherche. The practical part documents the current development of a new cell, on which tests were performed. Here is an overview of the influence of gases and insulating materials on the resulting data, especially distortion and attenuation. Subsequently, a new design with improved hermetic insulation and sample displacement error correction is described. The principle of operation of the manufactured cell, including the control unit and software, is elaborated in the form of technical documentation. Finally, the functionality is verified by comparing the diffractograms of the powder standards and the graphite electrodes.