Development of the Beam Position Monitors for the Diagnostics of the Test Beam Line in the CTF3 at CERN

The work for this thesis is in line with the field of Instrumentation for Particle Accelerators,
so called Beam Diagnostics. It is presented the development of a series of
electro-mechanical devices called Inductive Pick-Ups (IPU) for Beam Position Monitoring
(BPM). A full set of 17 BPM units (16 + 1 spare), named BPS units, were built and
installed into the Test Beam Line (TBL), an electron beam decelerator, of the 3rd CLIC
Test Facility (CTF3) at CERN ¿European Organization for the Nuclear Research¿.
The CTF3, built at CERN by an international collaboration, was meant to demonstrate
the technical feasibility of the key concepts for CLIC ¿Compact Linear Collider¿ as a
future linear collider based on the novel two-beam acceleration scheme, and in order to
achieve the next energy frontier for a lepton collider in theMulti-TeV scale. Modern particle
accelerators and in particular future colliders like CLIC requires an extreme alignment
and stabilization of the beam in order to enhance its quality, which rely heavily on a beam
based alignment techniques. Here the BPMs, like the BPS-IPU, play an important role
providing the beam position with precision and high resolution, besides a beam current
measurement in the case of the BPS, along the beam lines.
The BPS project carried out at IFIC was mainly developed in two phases: prototyping
and series production and test for the TBL.
In the first project phase two fully functional BPS prototypes were constructed, focusing
in this thesis work on the electronic design of the BPS on-board PCBs (Printed Circuit
Boards) which are based on transformers for the current sensing and beam position measurement.
Furthermore, it is described the monitor mechanical design with emphasis on
all the parts directly involved in its electromagnetic functioning, as a result of the coupling
of the EM fields generated by the beam with those parts. For that, it was studied
its operational parameters, according the TBL specifications, and it was also simulated a
new circuital model reproducing the BPS monitor frequency response for its operational
bandwidth (1kHz-100MHz). These prototypes were initially tested in the laboratories of
the BI-PI section¿Beam Instrumentation - Position and Intensity¿ at CERN.
In the second project phase the BPS monitor series, which were built based on the experience
acquired during the prototyping phase, the work was focused on the realization of
the characterization tests to measure the main operational parameters of each series monitor,
for which it was designed and constructed two test benches with different purposes
and frequency regions. The first one is designed to work in the low frequency region,
between 1kHz-100MHz, in the time scale of the electron beam pulse with a repetition
period of 1s and an approximate duration of 140ns. This kind of test setups called Wire
Test-bench are commonly used in the accelerators instrumentation field in order to determine
the characteristic parameters of a BPM (or pick-up) like its linearity and precision
in the position measurement, and also its frequency response (bandwidth). This is done
by emulating a low current intensity beam with a stretched wire carrying a current signals which can be precisely positioned with respect the device under test. This test bench was
specifically made for the BPS monitor and conceived to perform the measurement data
acquisition in an automated way, managing the measurement equipment and the wire positioning
motors controller from a PC workstation. Each one of the BPS monitors series
were characterized by using this system at the IFIC labs, and the test results and analysis
are presented in this work.
On the other hand, the high frequency tests, above the X band in the microwave spectrum
and at the time scale of the micro-bunch pulses with a bunching period of 83ps
(12GHz) inside a long 140ns pulse, were performed in order to measure the longitudinal
impedance of the BPS monitor. This must be low enough in order to minimize the
perturbations on the beam produced at crossing the monitor, which affects to its stability
during the propagation along the line. For that, it was built the high frequency test bench
as a coaxial waveguide structure of 24mm diameter matched at 50¿ and with a bandwidth
from 18MHz to 30GHz, which was previously simulated, and having room in the
middle to place the BPS as the device under test. This high frequency test bench is able
to reproduce the TEM (Transversal Electro-Magnetic) propagative modes corresponding
to an ultra-relativistic electron beam of 12GHz bunching frequency, so that the Scattering
parameters can be measured to obtain the longitudinal impedance of the BPS in the
frequency range of interest.
Finally, it is also presented the results of the beam test made in the TBL line, with
beam currents from 3.5A to 13A (max. available at the moment of the test). In order
to determine the minimum resolution attainable by a BPS monitor in the measurement
of the beam position, being the device figure of merit, with a resolution goal of 5¿m at
maximum beam current of 28A according to the TBL specifications. / García Garrigós, JJ. (2013). Development of the Beam Position Monitors for the Diagnostics of the Test Beam Line in the CTF3 at CERN [Tesis doctoral no publicada]. Universitat Politècnica de València. https://doi.org/10.4995/Thesis/10251/34327 / TESIS

Identiferoai:union.ndltd.org:upv.es/oai:riunet.upv.es:10251/34327
Date05 December 2013
CreatorsGarcía Garrigós, Juan José
ContributorsFaus Golfe, Ángeles, Mora Mas, Francisco José, Universitat Politècnica de València. Departamento de Ingeniería Electrónica - Departament d'Enginyeria Electrònica
PublisherUniversitat Politècnica de València
Source SetsUniversitat Politècnica de València
LanguageEnglish
Detected LanguageEnglish
Typeinfo:eu-repo/semantics/doctoralThesis, info:eu-repo/semantics/acceptedVersion
SourceRiunet
Rightshttp://rightsstatements.org/vocab/InC/1.0/, info:eu-repo/semantics/openAccess

Page generated in 0.0032 seconds