Análise da metodologia de calibração dos detectores de NaI(Tl) do Laboratório de Monitoração in vivo do IPEN pelo método de Monte de Carlo / Analysis of the calibration methods of NaI(Tl) detectors at the in vivo monitoring laboratory of ipen using the Monte Carlo Method

Esse trabalho avalia a metodologia utilizada pelo Laboratório de Monitoração in vivo (LMIV) do IPEN na calibração dos detectores de NaI(Tl) pelo código PENELOPE- penEasy, baseado no Método de Monte Carlo. Inicialmente, foram realizadas simulações sem o simulador antropomórfico com o objetivo de encontrar uma geometria que representasse adequadamente os detectores para exames de corpo inteiro e de tireoide. A aquisição dos dados experimentais foi realizada pelo software GENIE, que é utilizado na rotina do laboratório e as áreas dos fotopicos foram calculadas pelos softwares GENIE e GNUPLOT e pelo método descrito no Ciemat Technical Report, que também fornece os valores de eficiência de fotopico. Fontes de 60Co, 137Cs, 152Eu, 207Bi e 241Am calibradas foram utilizadas nas medições. Os resultados mostraram que o software GENIE é apropriado para ser utilizado nas calibrações de rotina e fontes que possuam picos isolados e definidos pelo detector de NaI(Tl), são ideais para serem utilizadas nas calibrações em eficiência. A simulação MC indica que é possível substituir calibrações experimentais por simulações utilizando uma geometria simplificada do detector de NaI(Tl), mas as grandezas relevantes como a fonte, o detector e o sistema eletrônico devem ser conhecidos com detalhes para que as fontes de erros sistemáticos sejam as menores possíveis. / The calibration methods of NaI(Tl) detectors of the in vivo Monitoring Laboratory (LMIV) of IPEN were analyzed by comparing experimental results with simulations performed with the PENELOPE- penEasy Monte Carlo radiation program. The first simulations were performed without the anthropomorphic simulator aiming to find a simplified detector geometry that could represent the whole body and thyroid detectors. The data acquisition was performed with the GENIE software, which is used in the laboratory routine, and the photopeak areas were calculated with the GENIE and GNUPLOT softwares as well as using the method described on the Ciemat Technical Report. Standard sources of 60Co, 137Cs, 152Eu, 207Bi and 241Am were used in the measurements. The results show that GENIE is a suitable software to be used in the laboratory routine and ideal sources are those that present isolated and well-defined photopeaks in the spectra measured with NaI(Tl) detectors. The Monte Carlo study reveals that the simplified models of the NaI detectors are able to provide a full- energy peak efficiency curve but it is necessary to know all details of the significant quantities as the source, the detector and the electronics in order to minimize sources of systematic errors.

Identiferoai:union.ndltd.org:usp.br/oai:teses.usp.br:tde-10022014-105712
Date11 December 2013
CreatorsKakoi, Adélia Aparecida Yuka
ContributorsMoralles, Mauricio
PublisherBiblioteca Digitais de Teses e Dissertações da USP
Source SetsUniversidade de São Paulo
LanguagePortuguese
Detected LanguagePortuguese
TypeDissertação de Mestrado
Formatapplication/pdf
RightsLiberar o conteúdo para acesso público.

Page generated in 0.0023 seconds