It was demonstrated in recent studies that Complementary Limb Motion Estimation (CLME) is robust approach for controlling active knee prostheses. A wearable sensor system is then needed to provide inputs to the controller in a real-time platform. In the present work, a wearable sensor system based on magnetic and inertial measurement units (MIMU) together with a simple calibration procedure were proposed. This sensor system was intended to substitute and extend the capabilities of a previous device based on potentiometers and gyroscopes. The proposed sensor system and calibration were validated with an Optical Tracking System (OTS) in a standard gait lab and first results showed that the proposed solution had a performance comparable to similar studies in the literature.
Identifer | oai:union.ndltd.org:UPSALLA1/oai:DiVA.org:liu-81244 |
Date | January 2012 |
Creators | Almeida, Eduardo Carlos Venancio de |
Publisher | Linköpings universitet, Biomedicinsk instrumentteknik, Linköpings universitet, Tekniska högskolan |
Source Sets | DiVA Archive at Upsalla University |
Language | English |
Detected Language | English |
Type | Student thesis, info:eu-repo/semantics/bachelorThesis, text |
Format | application/pdf |
Rights | info:eu-repo/semantics/openAccess |
Page generated in 0.0024 seconds