The Leksell Gamma Knife® (LGK) is a device for performing radiosurgery. The LGK contains approximately 200 radioactive sources whose beams intersect in a focal point in order to treat brain tumours. Quality assurance tools are used at Elekta to indirectly assess the number of sources in an LGK from the total amount of radiation. In order to increase patient safety, regulatory agencies have been asking for evidential proof of the number of sources in the LGK. This thesis' goal is to directly measure each source in the LGK and optimize the total detection time. To do this, a source detection system was developed with two parts, a radiation detection system and a moving gantry. Initial tests of the design were performed at Elekta and a final test was performed on an LGK at Karolinska Universitetssjukhuset. The results show that the proposed design has the possibility of detecting all sources in an LGK.
Identifer | oai:union.ndltd.org:UPSALLA1/oai:DiVA.org:kth-279067 |
Date | January 2020 |
Creators | Andersson, Simon |
Publisher | KTH, Skolan för kemi, bioteknologi och hälsa (CBH) |
Source Sets | DiVA Archive at Upsalla University |
Language | English |
Detected Language | English |
Type | Student thesis, info:eu-repo/semantics/bachelorThesis, text |
Format | application/pdf |
Rights | info:eu-repo/semantics/openAccess |
Relation | TRITA-CBH-GRU ; 2020:111 |
Page generated in 0.0016 seconds