The first part of thesis is theoretical and it is the introduction into conductometry. It defines the essential terms of this field as well as contact and contactless methods of measurement of fluid conductivity and also the influential parasitic phenomena and possibilities of their elimination. The second part is dealing with design of the contactless transformer conductivity probe. The characteristics of designed probe are verified in the practical part of the thesis. The final part of the thesis is dealing with measurement according to the defined points in the assignment. The characteristics of several contact probes are verified and also the characteristic of the contactless transformer probe, which was implemented according to the design from the second part. Outcomes of these measurements establish the suitable power supply frequencies and currents for all these probes.