The diploma thesis is aimed at the constructional design of diaphragm pumps with a linear motor. This pump is supposed to be used in medicine for pumping the blood or as an artificial heart. The basic facts, which are focused on the function of the heart and artificial heart, are mentioned in the first parts of the thesis. The next parts describe the diaphragm pumps and the constructional ways of their diaphragm fixation. Furthermore, the basic calculations, describing the function of the pump, are derived there. The basic parameters of the pump are also calculated. The thesis also contains the description of the body pump construction, whose working room is optimized by CFD calculation before the pump is manufactured. The final part of the thesis is concentrated on the measurement evaluation of produced pump model.