Microbubble ultrasound (US) contrast agents have been used to monitor the progression of anti-angiogenic chemotherapies. However, US backscatter measurements used in contrast imaging are inherently variable, given the presence of many microbubbles of random position and size. A model was developed to investigate the influence of US scanner and microbubble characteristics on these variable measurements. The Coefficient of Variation was used to measure variability. It was found that an optimum excitation frequency exists that minimizes this variability. In the case of DefinityTM, a 2.25 MHz centre-frequency pulse yielded a less variable measurement than at 5 MHz. Conversely, decreasing microbubbble concentration was found to significantly increase variability. Evidence suggests that microbubbles are no longer Rayleigh scatterers at sufficient low concentrations. Post-processing was found to aid in reducing measurement variability by averaging samples where microbubble positions are uncorrelated. As well, reduction can be achieved by averaging about a region-of-interest of uniform perfusion.
Identifer | oai:union.ndltd.org:TORONTO/oai:tspace.library.utoronto.ca:1807/33547 |
Date | 27 November 2012 |
Creators | Sureshkumar, Ahthavan R. |
Contributors | Burns, Peter N. |
Source Sets | University of Toronto |
Language | en_ca |
Detected Language | English |
Type | Thesis |
Page generated in 0.0017 seconds