Avaliação comparativa da dimensão do sulco gengival por tomografia de coerência óptica e sondas periodontais em indivíduos saudáveis

Submitted by Fabio Sobreira Campos da Costa (fabio.sobreira@ufpe.br) on 2017-02-14T13:32:22Z
No. of bitstreams: 2
license_rdf: 1232 bytes, checksum: 66e71c371cc565284e70f40736c94386 (MD5)
Dissertacao LUANA OSÓRIO FERNANDES.pdf: 1894600 bytes, checksum: 262777160a929e18e667f73cb8233ae7 (MD5) / Made available in DSpace on 2017-02-14T13:32:22Z (GMT). No. of bitstreams: 2
license_rdf: 1232 bytes, checksum: 66e71c371cc565284e70f40736c94386 (MD5)
Dissertacao LUANA OSÓRIO FERNANDES.pdf: 1894600 bytes, checksum: 262777160a929e18e667f73cb8233ae7 (MD5)
Previous issue date: 2015-02-11 / Este estudo avaliouo desempenho daTomografia de Coerência
Ópticaem1325nmcomoum método não invasivopara medir a profundidadedo
sulco gengivalde indivíduossaudáveisem comparação
comsondasperiodontaistradicionais emumambiente clínico.Dois
observadoresavaliaram aprofundidade dosulco gengivalde 445sitios emdentes
anteriores, tanto superior einferior,de 23 indivíduos saudáveis, por meio de três
ferramentas: sonda periodontal Carolinado Norte (NC), sonda
computadorizadaFloridaProbe (FP)eTomografia por Coerência Óptica(OCT).
Foi obtido oíndice de refraçãogengival,necessário para correção dos valores
obtidos por OCT,a partir da análisede19 amostras provenientes de cirurgia
periodontal. Para cada técnica, foiregistradoo grau de desconforto/dor, através
da escala verbal de 4 pontos(VRS-4) e Escala Numérica da Dor (END), e o
tempo com auxílio de cronômetro digital.As imagens de OCT
mostraramestruturas anatômicasclinicamente relevantes do ponto de vista
periodontalnaáreaavaliada.O valor médiocalculadopara o índicede refraçãoda
gengivafoi de 1,41±0,06, dado utilizado para realizar a correção dos valores
obtidos pelo OCT, através de regra de três simples.As imagens de OCT
proporcionaram analisar o correspondente ao sulco gengivalhistológicocom
valor médiode 0,86±0,27 milímetros.A sondaNCresultou em maiores valores de
profundidade(1,43 ±0,56 milímetros), seguido pelaFP(1,25±0,58 milímetros)
que proporcionou umamaior frequência dedesconforto edor.O tempo registrado
para cada técnica foi maior para o OCT (19,61±8,98 minutos) decorrente do
caráter laboratorial do equipamento, seguido da FP (2,44±1,04 minutos) e NC
(1,77±0,40 minutos), ambas adaptadas para aplicação clínica.Diante dos
resultados, pode-se concluir que o OCTéummétodo promissorparaa análise in
vivodasaúde periodontal, permitindoo diagnósticoprecoce e preciso, de
maneiranão invasiva, e que sua influência e interferênciasobre a
avaliaçãoclínicaéinexistente, diferente dousodeumasonda periodontal.Porém,
ainda se faz necessário a adaptação do equipamento para utilização na clinica
odontológica. / This study evaluatedthe performance ofOpticalCoherenceTomographyin
1325nmas a noninvasivemethodto measure the depthof the gingival sulcusof
healthysubjects compared totraditionalperiodontalprobesin a clinical
setting.Twoobservers evaluatedthe depth of thegingival sulcusof 445sitesin
anterior teeth, both top and bottomof 23individuals, using threetools: periodontal
probeNorth Carolina (NC), electronic probeFloridaProbe (FP) and Optical
CoherenceTomography(OCT). Was obtainedgingivalindex of refractionneeded
tocorrect thevalues obtained byOCT,from the analysisof 19samples
fromperiodontalsurgery. For eachtechnique,the degree of discomfort/painwas
recordedbyverbal4-point scale(VRS-4) and NumericalPainScale (NPS),
andtimeusing a digitalstopwatch.The OCTimages showedclinically
relevantanatomical structures of theperiodontalviewpoint in thearea
evaluated.Theaverage value calculated forthegumrefractive indexwas 1.41±
0.06, as used to perform thecorrection of values obtainedbyOCT,throughsimple
rule of three.TheOCTimagesprovidedanalyze
thehistologicalsulcuscorresponding toan average valueof 0.86±0.27 mm.The
CNproberesulted in greaterdepth values(1.43± 0.56millimeters), followed by
FP(1.25±0.58 mm) which gavea higher frequency ofdiscomfort and
pain.Thetime recordedfor each techniquewas higher fortheOCT(19.61 ±8.98
minutes) due to thecharacter oflaboratoryequipment, followed by FP(2.44±1.04
minutes) andNC (1.77 ±0, 40 minutes), bothadaptedfor clinical application. With
the results, we can concludethatOCTis apromising method forin vivoanalysisof
periodontalhealth, enabling early and accurate diagnosis, noninvasively,
andthat its influenceandinterferenceon clinical evaluationis lacking,differentthe
use ofa periodontalprobe.However,the upgrading of equipmentis still neededfor
usein the dentalclinic.

Identiferoai:union.ndltd.org:IBICT/oai:repositorio.ufpe.br:123456789/18317
Date11 February 2015
CreatorsFERNANDES, Luana Osório
ContributorsGOMES, Anderson Stevens L.
PublisherUniversidade Federal de Pernambuco, Programa de Pos Graduacao em Odontologia, UFPE, Brasil
Source SetsIBICT Brazilian ETDs
LanguageBreton