In our work we have studied the ultrasonic imaging systems and their use in ophthalmology, especially with the device Nidek 4000. We described ophthalmological examination methods. In addition, we are using the simulation program Field II. It simulated eye phantom and created his B-scan and biometry, where we compared the effects of different central frequency ultrasonic probes and different speeds of sound in the resulting values. We also created phantoms using agarose gel and materials of different properties. On phantoms, we studied the effect of the velocity of ultrasound in measurement results, effect of the concentration of the agarose gel to the velocity of sound. And we created phantoms simulating the human eye. Measurement protocol was created for use in teaching.