This thesis deals with the software development to calibration and interpretation of image data measured by a LEEM device. As the imaging technique is uniquely suited for in-situ studies of surface dynamical processes, the attention is mainly paid to methods enabling the evaluation of measurement time series. The phase correlation method based on Fourier transform of images is proposed to temperature shift correction between consecutive frames. The thesis describes the methods of additive and impulse noise filtering, image visualization, the filtration of secondary electrons and the determination of I-V curves from measured image data. Implemented methods are described not only in terms of their mathematical origin, but also with emphasis on the revealing of critical aspects associated with their use. The thesis also focuses on the application of the created algorithm to image data capturing the spatial and temporal evolution of 4,4’-biphenyl-dicarboxylic acid surface phases induced by sample annealing. Based on these evaluations, a suitable procedure is then determined to perform accurate detection and compensation of temperature shift said.