External beam radiation therapy for prostate cancer can result in urinary, sexual, and rectal side effects, often impairing quality of life. A polyethylene glycol-based product, SpaceOAR hydrogel (SOH), implanted into the connective tissue between prostate gland and rectum can significantly reduce the dose received by the rectum and hence risk of rectal toxicity. The optimal way to manage the hydrogel and rectal structures for plan optimization is therefore of interest.
A retrospective planning study was completed with 13 patients to examine optimal
planning and treatment methods. Computerized tomography (CT) scans were taken
pre- and post-SOH implant. Six hypofractionated (60 Gy in 20 fractions) treatment
plans were produced per patient using either a structure of rectum plus the hydrogel,
termed composite rectum wall (CRW), or rectal wall (RW) as the inverse optimization
structure and intensity modulated radiotherapy (IMRT) or volumetric modulated arc
therapy (VMAT) as the treatment technique. Dose-volume histogram metrics were
compared between plans to determine which optimization structure and treatment
technique offered the maximum rectal dose sparing. RW structures offered a statistically
significant decrease in rectal dose over CRW structures, whereas the treatment
technique (IMRT vs VMAT) did not significantly affect the rectal dose. However,
there was improvement seen in bladder and penile bulb dose when VMAT was used
as a treatment technique over IMRT. Overall, treatment plans using the RW optimization
structure offered the lowest rectal dose while VMAT treatment technique offered the lowest bladder and penile bulb dose. These treatment techniques and optimization structures have now been implemented at BC Cancer - Victoria based on this retrospective study.
SOH implant has been shown not to be equally effective in all patients. Determining
a priori patients in which the implant will offer most benefit allows for effective
management of SOH resources. Several factors have been shown to be correlated to
reduction in rectal dose including distance between rectum and planning treatment
volume (PTV), volume of rectum in the PTV and change in rectum volume pre- to
post-SOH. Several of these factors along with other pre-SOH CT metrics were found
via multiple linear regression models to predict reduction of rectal dose using data from 21 patients who received SOH implant. Two high rectal dose metrics were modeled, change in the relative volume receiving 55 Gy and change in the partial high dose integral, integrating over the dose-volume histogram (DVH) from 55 Gy to 60 Gy. Models were also produced to predict pre-SOH RV55Gy. These models offered R-squared between 0.57 and 0.87 with statistical significance in each model. Applying a 3.5% lower limit on pre-SOH RV55Gy removed one third of patients as implant candidates. This may offer a clinically useful tool in deciding which patients should receive SOH implant given limited resources. Predictive models, nomograms and a workflow diagram were produced for clinical management of SOH implant. / Graduate
Identifer | oai:union.ndltd.org:uvic.ca/oai:dspace.library.uvic.ca:1828/11200 |
Date | 27 September 2019 |
Creators | Paetkau, D. Owen |
Contributors | Bazalova-Carter, Magdalena, Gagne, Isabelle Marie |
Source Sets | University of Victoria |
Language | English, English |
Detected Language | English |
Type | Thesis |
Format | application/pdf |
Rights | Available to the World Wide Web |
Page generated in 0.002 seconds