Caracterização de dosímetros de alanina/RPE para irradiadores de raios-X de baixa energia

Submitted by Fabio Sobreira Campos da Costa (fabio.sobreira@ufpe.br) on 2017-07-14T13:38:53Z
No. of bitstreams: 2
license_rdf: 811 bytes, checksum: e39d27027a6cc9cb039ad269a5db8e34 (MD5)
Dissertação (versão final) Leoncio.pdf: 3177730 bytes, checksum: 3ae8af6eb5ff1a7f234df372ba6107db (MD5) / Made available in DSpace on 2017-07-14T13:38:53Z (GMT). No. of bitstreams: 2
license_rdf: 811 bytes, checksum: e39d27027a6cc9cb039ad269a5db8e34 (MD5)
Dissertação (versão final) Leoncio.pdf: 3177730 bytes, checksum: 3ae8af6eb5ff1a7f234df372ba6107db (MD5)
Previous issue date: 2016-06-30 / FACEPE / Nos últimos anos, a crescente dificuldade para aquisição e transporte internacional de
irradiadores de raios gama, como o irradiador de Co-60 GammaCell 220 da Nordion, levou ao
desenvolvimento de irradiadores de altas taxas de dose com base em feixes de raios-X, que
operam na faixa de 150kV. Uma das aplicações deste tipo de irradiador é a irradiação de
insetos, de amostras de sangue, etc. Para estes procedimentos, é importante que se tenha
dosímetros que possam monitorar os processos de irradiação. O objetivo deste trabalho é o
estudo e caracterização da alanina para que possa ser aplicada em dosimetria de irradiadores
de raios-X de altas taxas de dose. A alanina é um aminoácido com número atômico efetivo
próximo ao do tecido humano e é utilizada para dosimetria por ressonância paramagnética
eletrônica (RPE) para aplicações em radioterapia. Neste trabalho, foram produzidas pastilhas
de alanina com diferentes aglutinantes, bem como pastilhas puras revestidas com duas
camadas de verniz, para serem utilizadas como dosímetros. Foram utilizados como
aglutinantes parafina, teflon e estearina em concentrações de 1 %, 2 % e 5 % da massa total
das pastilhas. Testes mecânicos, como ultrassom e rot-up, foram realizados para avaliar a
resistência mecânica das pastilhas produzidas. Após os testes mecânicos, verificou-se que as
pastilhas preparadas com alanina pura apresentavam resistência mecânica suficiente para
serem utilizadas como dosímetros. A reprodutibilidade de sua resposta RPE foi avaliada a
partir da irradiação de dez dosímetros com dose de 100 Gy com uma fonte de Co-60. Para
investigar a estabilidade da resposta da alanina-RPE com o tempo, três dosímetros foram
irradiados com uma dose de 200 Gy com uma fonte de Co-60, sendo a leitura realizada após
cinco horas da irradiação e releituras após 15, 40, 87, 110 e 137 dias da irradiação. Para obter
as curvas de calibração para raios gama e raios – X, amostras foram irradiadas,
respectivamente, com doses de 0 a 1 kGy e de 18,7 Gy a 206 Gy. Os resultados mostram que
a resposta dos dosímetros de alanina-RPE apresentou reprodutibilidade com um coeficiente de
variação de 1,2%. O estudo do desvanecimento da resposta dos dosímetros mostrou uma
variação de 0,8% em 137 dias. De acordo com os resultados obtidos, o sistema de dosimetria
alanina-RPE pode ser utilizado, de forma confiável, como um sistema padrão de transferência
para as aplicações do irradiador de raios-X de baixa energia, o RS 2400. / In recent years, the increasing difficulty in acquisition and international transport of gamma
ray radiators, as irradiator Co-60 Gammacell 220 of Nordion led to the development of high
dose rate irradiators based on X-ray beams that they are operating in the 150kV range. One
application of this type of irradiator and irradiation of insects, blood samples, etc. For these
procedures, it is important to have dosimeters that can monitor the irradiation processes. The
objective of this work is the study and characterization of alanine that can be applied in
dosimetry irradiating X-ray of high dose rates. Alanine is an amino acid in effective atomic
number close to that of human tissue and is used for dosimetry for electron paramagnetic
resonance (EPR) for applications in radiotherapy. In this work, alanine pellets were produced
with different binders as well as pure tablets coated with two coats of varnish to be used as
dosimeters. They were used as binders paraffin, teflon and stearin in concentrations of 1%,
2% and 5% of the total mass of the pastilles. Mechanical tests, such as ultrasound and rot-up
were conducted to evaluate the mechanical strength of the tablets produced. After mechanical
testing, it was found that the tablets prepared with pure alanine had sufficient mechanical
strength to be used as dosimeters. The reproducibility of EPR response was evaluated from
the irradiation ten dosimeters with dose 100 Gy with a source of Co-60. To investigate the
stability of alanine EPR response over time of three dosimeters were irradiated with a dose of
200 Gy with a source of Co-60, the reading being performed after five hours of irradiation and
readings after 15, 40, 87, 110 and 137 days of irradiation. To obtain the calibration curves for
gamma rays and X - ray irradiated samples were, respectively, at doses from 0 to 1 kGy and
18.7 Gy to 206 Gy. The results show that the response of EPR-alanine dosimeters showed
reproducibility with a coefficient of variation 1.2%. The study fading dosimeters response
showed a variation of 0.8% in 137 days. According to the results, the system alanine EPR
dosimetry can be used reliably as a pattern transferring system for irradiating the applications
of low-energy X-rays, the RS 2400.

Identiferoai:union.ndltd.org:IBICT/oai:repositorio.ufpe.br:123456789/19638
Date30 June 2016
CreatorsSILVA NETO, Leoncio de Barros e
Contributorshttp://lattes.cnpq.br/5586454739210184, BARROS, Vinícius Saito Monteiro de, ASFORA, Viviane Khoury
PublisherUniversidade Federal de Pernambuco, Programa de Pos Graduacao em Tecnologias Energeticas e Nuclear, UFPE, Brasil
Source SetsIBICT Brazilian ETDs
LanguagePortuguese
Detected LanguagePortuguese
Typeinfo:eu-repo/semantics/publishedVersion, info:eu-repo/semantics/masterThesis
Sourcereponame:Repositório Institucional da UFPE, instname:Universidade Federal de Pernambuco, instacron:UFPE
RightsAttribution-NonCommercial-NoDerivs 3.0 Brazil, http://creativecommons.org/licenses/by-nc-nd/3.0/br/, info:eu-repo/semantics/openAccess

Page generated in 0.0024 seconds