This work covers an area of heavy metals detection by miniaturized thick film sensors. Working electrode of these sensors was fabricated using thick film paste that was prepared as a nanocomposition of carbon nanotubes powder, suitable vehicle and a thinner. The response of fabricated electrodes to cadmium and lead ions was measured using differential pulse voltammetry in an acetate buffer solution using a three electrode sytem. The results were compared with the carbon electrode made from special commercial thick film polymer paste. Finally the possibility of working electrode surface activation with nitric acid and heavy metals on real biological sample detection is described here.